Developed by SAMx, the PCIMax imaging board interfaced the SAMx software HiMax and MaxView to the SEM.- Signals· X and Y programmable scan from 0.1 Volt to 10 Volt. · External control signal to the SEM ( · 12, · 5, 0, -5 et -12 Volt ). · Programmable blanking signal, 0 or 5 Volt. · Programmable filament shutdown signal. · Two video input signals (SE, BSE for example). · 15 x-ray input signals ED/WD. - Image· Size from 64 to 2048 pixels per line. · TV rate at 256*256 resolution. · X-ray maps acquired at 10 us minimum dwell time. - Board/Connectors· Standard PCI board 170 mm long. · X-ray connector on the board. · Analog connector on front panel. - Communication with the PC. · Automatically recognized by the PC BIOS. · One interrupt (IRQ). · One DMA channel. |